XI.Matters needing attention and warnings of trocar

1. When using this laparoscopic trocar, aseptic operation specifications should be strictly followed;

2. Please carefully check the packaging of this laparoscopic trocar before use, if the blister packaging is damaged, please stop using it;

3. The laparoscopic trocar is sterilized by ethylene oxide, and.

4. This product is for one-time use and cannot be sterilized after use.

5. Please check whether the product is within the validity period before use. The sterilization validity period is three years, and the product beyond the validity period is strictly prohibited.

6. During abdominal surgery, if a proper pneumoperitoneum cannot be formed and maintained, the available operating space may be reduced, which will hinder the forward movement of the puncture cone and increase the risk of visceral tissue injury.

7. Only physicians with rich experience and familiarity with endoscopic technology can perform endoscopic surgery. Before surgery, doctors should consult relevant books and literature to understand the technology, complications and risks of endoscopy.

8. In-line piercing cones are sharper and safer. Therefore, the force required during the insertion process is smaller. Using excessive force may reduce the user's control of the insertion angle and depth of the puncture cone, and increase the risk of harming internal tissues.

9. Adhesion, abnormal anatomy or other obstacles may be prevented or delayed, which will cause the puncture cone to damage the internal organ structure during blind puncture. It is recommended to use open laparoscopic surgery. After adding sutures to the abdominal fascia, use this product Import and use.

10. Before and after removing the disposable puncture cannula from the abdominal cavity, check the hemostasis of the surgical site. Electric heating or manual sutures can be used to control bleeding. Depending on the doctor’s judgment, a laparotomy may be required.

11. Once you enter the abdominal cavity, you must be careful not to re-press the puncture cannula. If enough force is applied to the front end to move the cone forward, it will cause damage to the internal tissues.

/single-use-trocar-product/

12. When using endoscopic instruments and accessories produced by different manufacturers at the same time during endoscopic surgery, check the compatibility between the different instruments, and check the electrical insulation and good grounding.

13. If the puncture skin incision is 10mm or longer, the deep fascia should be closed to reduce the possibility of incisional hernia.

14. During abdominal laparoscopic surgery, the patient should be in a supine position with head down and feet high. When using the first puncture cannula, lift the lower abdominal wall with one hand, and operate the puncture cannula with the other hand, and insert the puncture cannula from the umbilical skin incision at a 45-degree angle toward the bladder.

15. Sterilized products should be stored in a cool, dry, clean, well-ventilated, and non-corrosive gas environment.

16. After successful puncture, before using the endoscopic instrument, apply medical lubricant on the surface of the endoscopic instrument or the sealing ring of the trocar to reduce resistance and make the operation more convenient and quicker.

17. See the label for the production date.

18. Explanation of graphics, symbols and abbreviations used in packaging and labels.